Table 1.
Measure | ACE | ACE+ | P-value |
---|---|---|---|
Transection time* | 5.7 s | 4.5 s | <0.001 |
Initial hemostasis | 127/130 | 95/96 | 0.639 |
Thermal damage** | 2.19 mm | 1.69 mm | <0.001 |
Survival hemostasis | 33/33 | 32/32 | 1.000 |
Adhesions | 13/33 | 0/32 | <0.001 |
Burst pressure | 1,208 mmhg | 1,514 mmhg | 0.002 |
Average visual obstruction | 4.57% | 2.72% | <0.001 |
Notes:
Adjusted for vessel size and device length via analysis of covariance (ANCOVA);
adjusted for device length and vessel type via ANCOVA.