Skip to main content
. 2013 Sep 16;6:151–154. doi: 10.2147/MDER.S51663

Table 1.

ACE® versus ACE®+ results summary

Measure ACE ACE+ P-value
Transection time* 5.7 s 4.5 s <0.001
Initial hemostasis 127/130 95/96 0.639
Thermal damage** 2.19 mm 1.69 mm <0.001
Survival hemostasis 33/33 32/32 1.000
Adhesions 13/33 0/32 <0.001
Burst pressure 1,208 mmhg 1,514 mmhg 0.002
Average visual obstruction 4.57% 2.72% <0.001

Notes:

*

Adjusted for vessel size and device length via analysis of covariance (ANCOVA);

**

adjusted for device length and vessel type via ANCOVA.