Table 3.

Bland-Altman analysis of the biometric measurements using swept source optical coherence tomography, partial coherence interferometry and immersion ultrasound A-scan biometry.

Parameter SS-OCT vs. PCI SS-OCT vs. IUS PCI vs. IUS
Mean±SD of Differences [μm] 95% Confidence Interval [μm] Mean±SD of Differences [μm] 95% Confidence Interval [μm] Mean±SD of Differences [μm] 95% Confidence Interval [μm]
AL 35±50 (p=0.005) −61 – 132 175±117 (p<0.001) −54 – 403 139±94 (p<0.001) −44 – 322
ACD 85±75 (p<0.001) −61 – 232 37±164 (p=0.325) −284 – 358 48±181 (p=0.248) −402 – 306
LT - - −268±236 (p<0.001) −731 – 195 - -
VD - - 415±155 (p<0.001) 111 – 719 - -

SS-OCT = swept source optical coherence tomography; PCI = partial coherence interferometry (IOL Master); IUS – immersion ultrasound A-scan biometry; AL = axial eye length; ACD = anterior chamber depth; LT = lens thickness; VD = vitreous depth; SD = standard deviation.