Table II.
Results of Polystyrene Standard Measurements with MFI4100, MFI5200, FlowCAM VS1, and FlowCAM PV
| Standard type | Specification | MFI4100 | MFI5200 | FlowCAM VS1 | FlowCAM PV |
|---|---|---|---|---|---|
| 5 μm count standard | 3,000 ± 300 part./mL (>3 μm)a | 2,906 ± 324 part./mLc | 3,203 ± 116 part./mLc | 2,779 ± 162 part./mLc | 2,974 ± 184 part./mLc |
| 2 μm size standard | 1.999 ± 0.020 μmb | 1.74 ± 0.28 μmd | 1.95 ± 0.35 μmd | 3.20 ± 1.39 μmd | 2.38 ± 0.90 μmd |
| 5 μm size standard | 4.993 ± 0.040 μmb | 5.10 ± 0.80 μmd | 5.12 ± 0.57 μmd | 5.94 ± 1.61 μmd | 4.66 ± 1.52 μmd |
| 10 μm size standard | 10.00 ± 0.08 μmb | 10.56 ± 1.22 μmd | 10.16 ± 1.16 μmd | 10.71 ± 2.41 μmd | 9.66 ± 1.43 μmd |
aBased on light obscuration
bBased on microscopy
cStandard deviation from three measurements
dFull peak width at half of the maximum height