Table 1. Raw and corrected (QuRE, ShoRAH) estimated error rates for the Roche's GS FLX Titanium XLR70, as estimated by shotgun-sequencing the plasmid TOPO2.1 with HCV/HIV-1 inserts, and mapping the reads against the TOPO2.1 reference using the QuRe read aligner/filter module.

        error prevalence (standard error) by type
data set (n = 70,000) mapped (n;%) error correction method overall error prevalence (standard error) insertions deletions mismatches homopolymeric non-homopolymeric
HCV 49,866; 71.24% non-corrected 1.26% (0.0096) 0.40% (0.0054) 0.39% (0.0054) 0.47% (0.0059) 1.33% (0.0168) 1.22% (0.0117)
    QuRe 0.19% (0.0037) 0.02% (0.0012) 0.14% (0.0032) 0.02% (0.0014) 0.27% (0.0076) 0.14% (0.0040)
    ShoRAH 0.18% (0.0037) 0.00% (0.0000) 0.12% (0.0031) 0.06% (0.0021) 0.24% (0.0073) 0.14% (0.0041)
HIV-1 62,303; 89.00% non-corrected 1.30% (0.0060) 0.36% (0.0032) 0.47% (0.0037) 0.47% (0.0036) 2.14% (0.0162) 1.05% (0.0062)
    QuRe 0.39% (0.0033) 0.06% (0.0013) 0.25% (0.0027) 0.08% (0.0015) 0.95% (0.0108) 0.23% (0.0029)
    ShoRAH 0.21% (0.0025) 0.00% (0.0001) 0.13% (0.0020) 0.07% (0.0015) 0.47% (0.0079) 0.13% (0.0022)