Table 1.

Measured parameters for phantom data

Distance
ITV (PET Edge)
ITV (40% threshold)
Traj 1 Traj 2 μ1 ± σ1 μ2 ± σ2 μ3 ± σ3 μ4 ± σ4 μ1 ± σ1 μ2 ± σ2 μ3 ± σ3 μ4 ± σ4
T1 76.2% 31.3% 1.62 ± 0.51 0.83 ± 0.21 0.82 ± 0.16 0.86 ± 0.15 1.83 ± 1.05 0.70 ± 0.10 0.65 ± 0.07 0.66 ± 0.04
T2 80.2% 31.3% 1.64 ± 0.46 0.88 ± 0.18 0.83 ± 0.16 0.87 ± 0.17 2.33 ± 1.40 0.75 ± 0.22 0.67 ± 0.10 0.68 ± 0.08
A1 89.0% 94.0% 1.79 ± 0.25 1.09 ± 0.12 1.01 ± 0.07 1.03 ± 0.05 2.06 ± 1.13 0.84 ± 0.18 0.84 ± 0.10 0.83 ± 0.05
A2 86.4% 60.0% 1.39 ± 0.19 0.93 ± 0.15 0.90 ± 0.11 0.95 ± 0.10 1.68 ± 0.88 0.67 ± 0.06 0.74 ± 0.06 0.77 ± 0.03

Abbreviations: ITV = internal target volumes; PET = positron emission tomography; Traj = trajectory.

Motion displacement recovery and measured to expected ITV measurements across the 4 gating algorithms for the phantom data. PET Edge was used to segment end-exhale and end-inhale images for superior-inferior distance measurements. Images at several signal-to-background ratios were used to collect the data for sphere sizes 2 to 4 cm (trajectory 1: n = 24 per algorithm; traj 2: n = 21 for A2, n=18 for rest). ITV measurements (mean and SD) are shown for all spheres (n = 6 per sphere size per algorithm).