Skip to main content
. Author manuscript; available in PMC: 2014 Aug 19.
Published in final edited form as: Chem Res Toxicol. 2013 Jul 24;26(8):1209–1217. doi: 10.1021/tx400121n

Figure 6.

Figure 6

Determination of the accuracy of the PheT analysis. A pooled smokers’ urine sample containing 2199 fmol/mL PheT was spiked with 6 increasing amounts of PheT and the analysis was carried out as described in Materials and Methods. Each point is the mean of 4 determinations. The added and determined amounts correlated, R2 = 1.00, and the accuracy was 95%. The y-intercept agrees with the amount of PheT in the pooled smokers’ urine sample.