. 2013 Oct 10;40(11):111907. doi: 10.1118/1.4823795

Table 2.

Technical settings for radiographic and CBCT imaging dose measurements. For the Pelvis AP and Head AP radiographic imaging protocols, the x-ray exposure was scaled for measurement by scaling the clinical x-ray exposure time. The clinical exposures and exposure times are given in brackets [].

  Pelvis Head Pelvis Standard-dose
  AP AP CBCT head CBCT
X-ray voltage (kVp) 80 100 125 100
X-ray current (mA) 200 200 80 20
X-ray milliseconds (ms) 1000 [50] 1000 [40] 13 20
Gantry rotation range (deg) Fixed Fixed 364 204
Number of projections 1 1 655 360
X-ray exposure (mAs) 200 [10] 200 [8] 681 144
Fan type Half Full Half Full
Bowtie filter Half Full Half Full
Blade X1 (cm) 6.8 13.6 6.8 13.6
Blade X2 (cm) 23.5 13.6 23.5 13.6
Blade Y1 (cm) 10.3 9.2 10.3 9.2
Blade Y2 (cm) 10.3 9.2 10.3 9.2