Skip to main content
. Author manuscript; available in PMC: 2014 Apr 1.
Published in final edited form as: J Magn Reson. 2013 Feb 8;229:10.1016/j.jmr.2013.02.003. doi: 10.1016/j.jmr.2013.02.003

Table 1.

Magnitude of chemical shift (in millimeters) associated with selection pulse with a given bandwidth at for a 100mm-wide selected region. The standard 180 degree pulses used in commercial sequences are typically 1kHz, spatial-spectral pulses are in the range of 3–5kHz and the most recent LASER pulses are in the range of 5–20kHz.

Frequency Water to Lipid Cho to NAA
Bw (kHz) 1.0 1.5 2.5 5.0 1.0 1.5 2.5 5.0
Shift for 3T
(mm)
42 28 17 8.5 15 10 7 3.5
Shift for 7T
(mm)
77 51 31 15.5 35 23 14 7