Table 1.
Condition | AoFm (L/min) | LVPsys (mm Hg) | LVPed (mm Hg) | LVEW (mm Hg · mL) | +dP/dt (mm Hg/s) | −dP/dt (mm Hg/s) | AoPm (mm Hg) | AoPsys (mm Hg) | AoPdia (mm Hg) |
---|---|---|---|---|---|---|---|---|---|
Baseline | 5.7 ± 1.5 | 91 ± 6 | 12 ± 0.4 | 6237 ± 109 | 791 ± 40 | −1187 ± 3 | 81 ±6 | 91 ±2 | 73 ±8 |
MVAD support | 6.3 ±1.7 | 70 ± 5 | 6 ±3.3 | 3299 ± 1212 | 628 ± 25 | −795 ± 138 | 85 ± 1 | 96 ±6 | 76 ±6 |
These data demonstrate unloading of native left ventricle and improvement in flow with miniaturized ventricular assist device support. AoFm, Mean aortic flow; LVPsys, peak (systolic) left ventricular pressure; LVPed, left ventricular end-diastolic pressure; LVEW, left ventricular external work; +dP/dt, positive rate of change in left ventricular pressure; −dP/dt, negative rate of change in left ventricular pressure; AoPm, mean aortic pressure; AoPsys, systolic aortic pressure; AoPdia, diastolic aortic pressure; MVAD, miniaturized ventricular assist device.