Table 3.

Parallelism assessment of IL-13 assay in human plasma under acidic conditions. The accuracy was calculated using the IL-13 concentration determined at dilution 1 : 2 as a reference value. Parallelism was considered acceptable when accuracy was between 70 and 130% for at least 5/6 samples at each dilution.

Dilution Raw result (pg/mL) Accuracy (O/E, %)
Observed (O) Expected (E)
Sample 1
1 : 2 0.425
1 : 3 0.250 0.283 88.2
1 : 4 0.195 0.213 91.8
1 : 6 N.C N.C. N.C.

Sample 2
1 : 2 1.280
1 : 3 0.760 0.853 89.1
1 : 4 0.525 0.640 82.0
1 : 6 0.362 0.427 84.8

Sample 3
1 : 2 0.745
1 : 3 0.517 0.497 104.0
1 : 4 0.425 0.373 114.1
1 : 6 0.285 0.248 114.8

Sample 4
1 : 2 1.190
1 : 3 0.763 0.793 96.2
1 : 4 0.540 0.595 90.8
1 : 6 0.365 0.397 92.0

Sample 5
1 : 2 1.330
1 : 3 1.020 0.887 115.0
1 : 4 0.830 0.665 124.8
1 : 6 0.553 0.443 124.8

Sample 6
1 : 2 0.270
1 : 3 0.207 0.180 114.8
1 : 4 0.148 0.135 109.3
1 : 6 N.C N.C. N.C.

N.C.: value not calculated as being outside the calibration range.