Table 2.
TruTip/epMotion and easyMAG extraction comparison from clinical specimens with two commercially-available real-time one-step RT-PCR master mixes.
| qScript™ One-Step qRT-PCR Kit | ABI TaqMan® Fast Virus 1step RT-PCR Master Mix |
||||||||
|---|---|---|---|---|---|---|---|---|---|
| Number of Samples |
easyMAG Ct range |
Mean ΔCta |
StDev | 95% confidence intervalb (CI) |
Number of Samples |
easyMAG Ct range |
Mean ΔCta |
StDev | 95% confidence intervalb (CI) |
| 4 | 17.00 – 20.30 | 1.42 | 0.92 | 1.42 ± 0.90 | Only 1 sample with Ct < 19.00 | ||||
| 30 | 20.31 – 23.60 | 1.82 | 1.02 | 1.82 ± 0.37 | 7 | 19.00 – 23.60 | 2.04 | 0.90 | 2.04 ± 0.67 |
| 74 | 23.61 – 26.90 | 2.14 | 0.98 | 2.14 ± 0.22 | 19 | 23.61 – 26.90 | 1.39 | 1.04 | 1.39 ± 0.47 |
| 40 | 26.91 – 30.20 | 2.40 | 0.65 | 2.40 ± 0.20 | 27 | 26.91 – 30.20 | 1.40 | 1.00 | 1.40 ± 0.38 |
| 12 | 30.21 – 33.50 | 1.60 | 0.86 | 1.60 ± 0.49 | 9 | 30.21 – 33.50 | 1.04 | 0.47 | 1.04 ± 0.31 |
| 5 | 33.51 – 36.80 | 1.57 | 1.23 | 1.57 ± 1.08 | 3 | 33.51 – 36.80 | 0.47 | 0.58 | 0.47 ± 0.66 |
| 3 | 36.81 – 40.10 | 0.97 | 1.62 | 0.97 ± 1.83 | 4 | 36.81 – 40.10 | −0.14 | 2.05 | −0.14 ± 2.01 |
| N = 168 | Average = | 1.89 | 1.03 | 1.89 ± 0.16 | N = 69 | Average = | 1.29 | 1.09 | 1.29 ± 0.26 |
Mean ΔCt calculated as the average TruTip/epMotion Ct value minus the average easyMAG Ct value.
95% CI values represent MeanΔCt ± CI