Distribution of urine samples using a workflow algorithm combining the Alfred 60/AST device, direct microscopy, and culture method

Test result No. of urine samples (total, 508)
Alfred 60 positive 142
    Microscopy positive c
    Culture positive 71
    Culture negative 71
Alfred 60 negative 366
    Microscopy positive 141
        Culture positive 8a
        Culture negative 133
    Microscopy negative 225
        Culture positive 1b
        Culture negative 224
a

False-negative Alfred 60/AST results correctly identified as positive by microscopy.

b

Represents 1 undiscovered positive culture result.