Table 1.

Pump performance comparison for of single-level TP, multi-level TP, and multi-level TP with double TP air chamber volume.

CD Design: Δt (min) ΔTC (deg) QT (μL/min deg) k (×10−2 / deg) Δ Vh (uL)
Single-level TP 5.6 13.6 1.04 1.24 75
Multi-level TP 8.3 14.3 0.67 1.11 78
Multi-level TP (with double TP air chamber volume) 3.7 12.0 1.81 1.59 80

Note: ΔVh is 80 μL, T0 is 25°C, Δt is the time required for liquid transfer, ΔTC is the change in CD surface temperature, QT is the flow rate per temperature gradient using Eqn. (2), k, is the model constant using Eqn. (3), ΔVh = VhV0, where Vh is obtained using Eqn. (5)