Table 1.

LC-MS/MS parameters performed for this study in the five core facilities.

Parameter TripleTOF 5600
#1
TripleTOF 5600
#2
LTQ-Orbitrap
Velos
Q Exactive Orbitrap Elite
LC cHiPLC Nanoflex cHiPLC Nanoflex UPLC UPLC nanoHPLC
Model Eksigent Eksigent 2D ultra
plus
Waters
nanoACQUITY
Waters
nanoACQUITY
Proxeon nEasy II
Peak width at half
height
17 sec variable <30 sec 6 sec variable
Trap Nano cHiPLC
Trap column 200
μm i.d. × 0.5 mm
Eksigent 3 um
(120Å) C18
Nano cHiPLC
Trap column 200
μm i.d. × 0.5 mm
Eksigent 3 um
(120Å) C18
Integrafrit,
100 μm i.d. ×
5cm of 5 μm
Magic C18AQ
Homemade
5 μm (200Å) C18
Homemade
5 μm (200Å)
Magic C18
Analytical column
model
Nano cHiPLC
column Eksigent
75 μm i.d.
Nano cHiPLC
column Eksigent
75 μm i.d.
Picofrit, 75 μm
i.d. 15 μm tip
Homemade 75
μm i.d.
Homemade 75
μm i.d.
Length 15 cm bed 15 cm bed 20 cm bed 25 cm bed 25 cm bed
Resin type ChromXP C18-CL
C18 on silica
ChromXP C18-CL
C18 on silica
Magic C18AQ
silica
Magic C18AQ
silica
Magic C18AQ
silica
Particle size 3 μm (120Å) 3 μm (120Å) 3 μm (100Å) 3 μm (100Å) 5 μm (100Å)
Average pressure
(psi)
1500 1800 2500 3500 1810
Sample injected Variable volume
for 200 ng
10 μL, about
< 1000 ng
15/35 μL about
< 1200 ng
4.5 μL, about
400 ng
12 μL, about
< 1200 ng
Gradient length for
data acquisition
120 min 120 min 90 min 90 min 90 min
LC flow rate 300 nL/min 300 nL/min 200 nL/min 300 nL/min 300 nL/min
Survey scan
resolution
~30,000 ~30,000 30,000 70,000 120,000
Survey scan m/z
range
400-1250 400-1250 395-1600 300-1750 400-1800
Survey scan length 250 msec 250 msec ~250 msec Max 120 msec 250 msec
Isotope exclusion ON No No Yes Yes No
Fragmentation
method
CID CID HCD HCD HCD
Charge for MS/MS +2 to +5 +2 to +5 +1 or higher +2 to +7 +2 or higher
Dynamic exclusion 10 sec after 1
occurrence
25 sec Yes, 60 sec
repeat count 1
30 sec 100 sec
Max. # of MS/MS
scans per period
20 50 10 12 10
MS/MS Scan Length 100 msec Variable for 2.8
sec period
Max ion time
250 msec
Max ion time
120 msec
250 msec
Parent ion selection
window
0.7 Da 0.7 Da 1.0 Da 1.2 Da 1.2 Da
MS/MS scan
resolution set
>19,000 ~15,000 7500 35,000 15,000
Maximum MS/MS
Fragment m/z
2000 2000 Variable 1750 Variable
Minimum MS/MS
Fragment m/z
100 65 100 100 100