Table 2.
Effect of temperature on equipment at 20 °C vs. 26 °C. One hundred and twenty-nine measurements were taken for each piece of equipment. Measurements from identical pieces of equipment were pooled for analysis
| Equipment | 20 °C | 26 °C | Paired difference | P value |
|---|---|---|---|---|
| Scope 1 | 34.30 ± 0.02 | 36.16 ± 0.07 | 1.86 ± 0.07 | <0.0001 |
| Scope 6 | 35.00 ± 0.07 | 37.05 ± 0.16 | 2.05 ± 0.17 | <0.0001 |
| Scope 3, 4 | 35.15 ± 1.13 | 36.90 ± 0.21 | 1.75 ± 1.32 | <0.0001 |
| Scope 2, 5 | 33.08 ± 0.79 | 35.02 ± 0.36 | 1.94 ± 0.46 | <0.0001 |
| Slide warmer 1 | 34.38 ± 0.16 | 36.38 ± 0.08 | 3.39 ± 0.20 | <0.0001 |
| Slide warmer 3 | 32.74 ± 0.36 | 35.89 ± 0.47 | 3.14 ± 0.55 | <0.0001 |
| Slide warmer 2, 4, 5 | 32.82 ± 0.29 | 35.76 ± 0.39 | 2.94 ± 0.48 | <0.0001 |
| Heating block 3 | 34.70 ± 0.93 | 35.19 ± 0.90 | 0.49 ± 1.21 | <0.0001 |
| Heating block 5 | 37.12 ± 0.08 | 37.67 ± 0.05 | 0.55 ± 0.09 | <0.0001 |
| Dry incubator 2 | 36.79 ± 0.17 | 37.7 ± 0.16 | 0.88 ± 0.27 | <0.0001 |
| Incubator 1 | 37.0 ± 0.08 | 37.0 ± 0 | 0.008 ± .09 | 0.32 |
| Incubator 11 | 36.60 ± 0.02 | 37.0 ± 0.009 | 0.10 ± 0.009 | <0.0001 |
| Incubator 12 | 36.4 ± 0 | 36.6 ± 0 | 0.20 ± 0.0 | <0.0001 |