. 2013 Aug 7;30(10):1389–1393. doi: 10.1007/s10815-013-0064-4

Table 2.

Effect of temperature on equipment at 20 °C vs. 26 °C. One hundred and twenty-nine measurements were taken for each piece of equipment. Measurements from identical pieces of equipment were pooled for analysis

Equipment 20 °C 26 °C Paired difference P value
Scope 1 34.30 ± 0.02 36.16 ± 0.07 1.86 ± 0.07 <0.0001
Scope 6 35.00 ± 0.07 37.05 ± 0.16 2.05 ± 0.17 <0.0001
Scope 3, 4 35.15 ± 1.13 36.90 ± 0.21 1.75 ± 1.32 <0.0001
Scope 2, 5 33.08 ± 0.79 35.02 ± 0.36 1.94 ± 0.46 <0.0001
Slide warmer 1 34.38 ± 0.16 36.38 ± 0.08 3.39 ± 0.20 <0.0001
Slide warmer 3 32.74 ± 0.36 35.89 ± 0.47 3.14 ± 0.55 <0.0001
Slide warmer 2, 4, 5 32.82 ± 0.29 35.76 ± 0.39 2.94 ± 0.48 <0.0001
Heating block 3 34.70 ± 0.93 35.19 ± 0.90 0.49 ± 1.21 <0.0001
Heating block 5 37.12 ± 0.08 37.67 ± 0.05 0.55 ± 0.09 <0.0001
Dry incubator 2 36.79 ± 0.17 37.7 ± 0.16 0.88 ± 0.27 <0.0001
Incubator 1 37.0 ± 0.08 37.0 ± 0 0.008 ± .09 0.32
Incubator 11 36.60 ± 0.02 37.0 ± 0.009 0.10 ± 0.009 <0.0001
Incubator 12 36.4 ± 0 36.6 ± 0 0.20 ± 0.0 <0.0001