Table 3.

The “trachea” like region of a variant of the COPDGene phantom was scanned in SS mode (80, 100, 120 and 140 kVp), in the DSDE mode (80/140Sn kVp, 140/80 kVp, and 100/140Sn kVp), and in the DSDE-SS mode (the same combinations as in the DSDE mode but with tube A or B set at min mAs). A consistent CTDIvol of 12 (+/− 0.1) mGy was used for each. The results here show the mean HU for a sampling of the analyzed parameter changes. Additional controlled scans comparing slice thickness (0.6 or 0.75mm), slice spacing (0.5 or 0.6mm), and scan direction (cranio-caudal or caudo-cranial) were also analyzed but not shown to have any meaningful difference in mean values compared to the control.

Phantom “Trachea” values
Scan Mode Mean Channels Pitch Rot. Time Kernel Tube Sn filtered
80 kVp SS −987 128 0.55 0.5 B35f A -
DSDE-SS −1002 64 0.55 0.5 B35f A -
DSDE-SS −1002 128 0.55 0.5 B35f A -
DSDE −1002 128 0.55 0.5 B35f A -
DSDE −1002 128 1.0 0.5 B35f A -
DSDE −1001 128 0.55 0.33 B35f A -
DSDE −1003 128 0.55 0.5 D30f A -
DSDE-SS −1003 64 0.55 0.5 B35f B -
DSDE-SS −1004 128 0.55 0.5 B35f B -
DSDE −1003 128 0.55 0.5 B35f B -
DSDE −1002 128 1.0 0.5 B35f B -
DSDE −1003 128 0.55 0.33 B35f B -
DSDE −1002 128 0.55 0.5 D30f B -

100 kVp SS −988 128 0.55 0.5 B35f A -
DSDE-SS −1004 64 0.55 0.5 B35f A -
DSDE-SS −1007 128 0.55 0.5 B35f A -
DSDE −1002 128 0.55 0.5 B35f A -

120 kVp SS −985 128 0.55 0.5 B35f A -

140 kVp SS −987 128 0.55 0.5 B35f A -
DSDE-SS −1002 64 0.55 0.5 B35f A -
DSDE-SS −1007 128 0.55 0.5 B35f A -
DSDE −1002 128 0.55 0.5 B35f A -
DSDE-SS −1000 64 0.55 0.5 B35f B yes
DSDE-SS −1001 128 0.55 0.5 B35f B yes
DSDE −1003 128 0.55 0.5 B35f B yes

Note: Reference CT number for internal “trachea” air is −1000 HU.