Skip to main content
. 2013 Aug 17;471(12):4047–4055. doi: 10.1007/s11999-013-3239-5

Table 2.

Translation (position) and orientation error of the ultrasound-based pedicle screw center relative to the CT-based pedicle screw center

Error type Healthy model, mean ± SD Degenerative model, mean ± SD
Translation R (mm) 0.16 ± 0.19 0.55 ± 0.59
Translation A (mm) −0.01 ± 1.22 −0.35 ± 0.40
Translation S (mm) 0.68 ± 0.38 1.28 ± 1.37
Rotation L-R (degrees) 1.92 ± 1.95 1.60 ± 1.56
Rotation P-A (degrees) −0.05 ± 0.42 0.81 ± 1.15
Rotation I-S (degrees) 0.40 ± 0.99 −0.79 ± 0.46