. 2013 Oct;26(5):266–279. doi: 10.1089/jamp.2012.1006

FIG. 6.

FIG. 6.

Comparison of experimental and CFD predictions of DF in the Optiflow nasal cannula and 20 cm of 10-mm diameter inlet tubing for a flow rate of 20 L/min and particle sizes of (a) 900 nm and (b) 4.7 μm. Particle trajectories are contoured according to velocity, and particle deposition locations are illustrated based on CFD predictions.