Table 6.
Landmarka | Galileos® (n = 10) | Next Generation i-CAT® (n = 10) | Wilcoxon rank-sum p-valueb | ||||||||||
Mean | Standard deviation | Median | Minimum | Maximum | Range | Mean | Standard deviation | Median | Minimum | Maximum | Range | ||
A | 1.82 | 2.0946 | 1.03 | 0.30 | 7.18 | 6.88 | 1.07 | 0.5954 | 0.86 | 0.36 | 2.00 | 1.64 | 0.6232 |
ANS | 1.07 | 0.3473 | 1.07 | 0.57 | 1.71 | 1.14 | 1.31 | 0.4274 | 1.39 | 0.59 | 1.91 | 1.32 | 0.1620 |
B point | 1.57 | 1.2072 | 1.31 | 0.28 | 4.73 | 4.44 | 1.12 | 0.4558 | 1.11 | 0.62 | 1.77 | 1.16 | 0.4723 |
Condylion | 0.98 | 0.2973 | 0.96 | 0.60 | 1.64 | 1.04 | 0.87 | 0.4691 | 0.80 | 0.17 | 1.71 | 1.54 | 0.3447 |
Gonion | 5.67 | 3.2599 | 5.54 | 0.55 | 11.48 | 10.94 | 5.90 | 2.5034 | 4.79 | 2.56 | 9.90 | 7.34 | 0.9698 |
Sella | 0.91 | 0.6333 | 0.64 | 0.30 | 2.28 | 1.98 | 0.76 | 0.2662 | 0.78 | 0.30 | 1.12 | 0.82 | 0.7337 |
Sigmoid notch | 0.77 | 0.4186 | 0.80 | 0.22 | 1.61 | 1.39 | 0.81 | 0.3583 | 0.79 | 0.22 | 1.41 | 1.18 | 0.7913 |
Lower molar | 1.14 | 0.4174 | 1.15 | 0.57 | 2.04 | 1.47 | 0.81 | 0.2695 | 0.77 | 0.42 | 1.14 | 0.72 | 0.0640 |
Md1 Root | 1.21 | 0.7345 | 0.87 | 0.36 | 2.27 | 1.91 | 1.48 | 0.7425 | 1.58 | 0.50 | 2.85 | 2.35 | 0.5205 |
Md1 tip | 0.77 | 0.2905 | 0.80 | 0.36 | 1.16 | 0.80 | 1.57 | 1.7454 | 0.88 | 0.47 | 4.93 | 4.46 | 0.6232 |
Me | 1.09 | 0.7998 | 0.92 | 0.20 | 2.91 | 2.71 | 0.79 | 0.4811 | 0.80 | 0.17 | 1.53 | 1.35 | 0.5205 |
Mx1 Root | 1.06 | 0.4346 | 0.94 | 0.59 | 1.94 | 1.35 | 1.15 | 0.4014 | 1.15 | 0.41 | 1.85 | 1.43 | 0.5205 |
Mx1 tip | 0.73 | 0.3170 | 0.67 | 0.14 | 1.24 | 1.10 | 0.88 | 0.4717 | 0.69 | 0.22 | 1.67 | 1.44 | 0.6232 |
Nasion | 2.91 | 1.9704 | 2.13 | 0.90 | 6.76 | 5.86 | 1.14 | 1.1828 | 0.85 | 0.22 | 4.27 | 4.04 | 0.0073 |
Orbitale | 1.86 | 1.3491 | 1.42 | 0.73 | 5.47 | 4.74 | 0.94 | 0.5225 | 0.80 | 0.37 | 2.22 | 1.84 | 0.0073 |
PNS | 1.19 | 0.6863 | 1.01 | 0.30 | 2.26 | 1.96 | 1.21 | 0.6705 | 0.97 | 0.32 | 2.28 | 1.97 | 0.9698 |
Pog | 2.10 | 2.0765 | 0.99 | 0.30 | 6.05 | 5.75 | 1.40 | 1.1435 | 1.13 | 0.30 | 4.22 | 3.92 | 0.6776 |
Upper molar | 0.86 | 0.2824 | 0.79 | 0.52 | 1.32 | 0.80 | 0.83 | 0.3922 | 0.90 | 0.00 | 1.45 | 1.45 | 0.8501 |
See Table 1 for definitions of the landmarks.
Significance probability associated with the Wilcoxon rank-sum test to determine whether the distributions of the Euclidean distance between dual measures were equal. Highlighting in bold indicates the significance value of <0.05.
Galileos is manufactured by Sirona Dental Systems Inc., Bensheim, Germany, and Next Generation i-CAT by Imaging Sciences International, Hatfield, PA.