Skip to main content
. 2013 Nov 15;8(11):e79185. doi: 10.1371/journal.pone.0079185

Table 1. Summary of the results obtained in terms of sensitivity and specificity for testing sera by immunofluorescence on cells transfected with AQP4-M23 or AQP4-M1 (AQP4-M1>M23) or AQP4-M1 (M1M23I).

Patient sera(n = 112) AQP4-M23 based test(M23 isoform) AQP4-M1 based test(M1>M23) AQP4-M1M23I based test(M1 isoform)
NMO (n = 40) Positive: 39 Positive: 11 Positive: 0
Sensitivity 39/40 = 97.5% (95% CI: 87%–100%) 11/40 = 27.5% (95% CI: 15%–44%) 0/40 = 0% (95% CI: 0%–9%)
Other diagnosis andhealthy donors (n = 72) Positive: 0 Positive: 0 Positive: 0
Specificity 72/72 = 100% (95% CI: 95%–100%) 72/72 = 100% (95% CI: 95%–100%) 72/72 = 100% (95% CI: 95%–100%)
Likelihood ratios LR+: ∞; LR−: 0.025 LR+: ∞; LR−: 0.725 LR+: ∞; LR−: 1

Note that the sensitivity is 97.5% and specificity is 100%.