Skip to main content
. 2012 May 25;39(6):3435–3444. doi: 10.1118/1.4718874

TABLE III.

Summary of measured Ergo performance characteristics.

  LEHR LEAP  
Native resolution (mm) at 10 cm 7.5 (7.4) 9.3 (10.6)  
Sensitivity (cps/MBq) 63.1 (59.7) 112.6 (113.8)  
(cpm/μCi) 140.1 (132.5) 249.9 (252.6)  
Pixel size (mm2) 3.28 × 3.18 (3.31 × 3.24)  
Uniformity <5% IU, <3% DU  
Count rate verified to be linear up to 5 Mcps  
Energy linearity linear in the range of 50–350 keV  

Note: LEHR = low energy high resolution; LEAP = low energy all purpose; IU = integral uniformity, DU = differential uniformity. The numbers in parentheses are the manufacturer specifications.