Skip to main content
. Author manuscript; available in PMC: 2014 Dec 1.
Published in final edited form as: Am J Ophthalmol. 2013 Sep 25;156(6):10.1016/j.ajo.2013.08.001. doi: 10.1016/j.ajo.2013.08.001

Table 4.

Frequency of statistically abnormal measurements as flagged on the retinal nerve fiber layer and Ganglion Cell Analysis printouts of Cirrus high-definition optical coherence tomography in the normal control subjects and glaucoma patients.

OCT parameters Normal group Glaucoma group Positive Likelihood Ratio Negative Likelihood Ratio
p<5% p<1% p<5% p<1% p<5% p<1% p<5% considered abnormal Only p<1% considered abnormal
Average RNFL 10 (11%) 8 (8.8%) 42 (71.2%) 25 (42.4%) 6.47 (3.53-11.88) 4.82 (2.33-9.96) 0.32 (0.22-0.49) 0.63 (0.50-0.79)
Average GC/IPL 10 (11%) 5 (5.5%) 37 (62.7%) 30 (50.9%) 5.71 (3.08-10.6) 9.25 (3.81-22.50) 0.42 (0.30-0.59) 0.52 (0.40-0.68)
RNFL quadrant 11(12.1%) 9 (9.9%) 55 (93.2%) 49 (83.1%) 7.71 (4.41-13.49) 8.39 (4.47-15.78) 0.08 (0.03-0.20) 0.18 (0.11-0.33)
RNFL sectors 27(29.7%) 4 (4.4%) 57 (96.6%) 49 (83.1%) 3.25 (2.37-4.48) 18.89 (7.2-49.60) 0.05 (0.01-0.19) 0.18 (0.10-0.31)
GC/IPL sectors 8 (8.8%) 6 (6.6%) 53 (89.8%) 49 (83.1%) 10.22 (5.24-19.92) 12.60 (5.76-27.53) 0.11 (0.05-0.24) 0.18 (0.10-0.32)
Min GC/IPL 7 (7.7%) 6 (6.6%) 50 (84.8%) 42 (71.2%) 11.02 (5.36-22.63) 10.80 (4.90-23.79) 0.16 (0.09-0.30) 0.30 (0.21-0.46)

OCT: Optical Coherence Tomography; RNFL: Retinal Nerve Fiber Layer; GC/IPL: Ganglion Cell/inner plexiform layer; Min: Minimum.