Skip to main content
. 2013 Dec;51(12):3960–3967. doi: 10.1128/JCM.01646-13

Table 3.

Standard curve parameters and amplification efficiencies for different plasmid virus cDNAs in PCRa

Plasmid virus Intercept parameters with:
Slope parameters with:
Efficiency parameters with:
FAM-RIp Cy5-ENp BOXTO FAM-RIp Cy5-ENp BOXTO FAM-RIp Cy5-ENp BOXTO
HRV-A1b 40.17 39.15 −3.43 −3.31 0.96 1.00
HRV-B14 38.92 41.25 −3.48 −3.35 0.94 0.99
HRV-A85 40.04 41.93 −3.37 −3.30 0.98 1.01
CVA16 39.97 39.82 −3.36 −3.32 0.98 1.00
CVB4 40.71 39.20 −3.33 −3.24 1.00 1.03
EV11 39.46 42.60 −3.33 −3.24 1.00 1.04
a

Linear regression curves (y = mx + b, where y = CT, m = slope, x = log10[copies/PCR], and b = intercept when x = 0) and efficiencies (E = 10−1/m − 1) were determined with 10-fold dilutions containing 5 × 107 to 5 plasmid copies/reaction.