. 2013 Dec;51(12):4001–4007. doi: 10.1128/JCM.02056-13

Table 5.

Differentiation of stx1 and stx2 in clinical specimens compared to bidirectional sequencing

Gene target(s) Resultsa
Performance (% [95% CI])b
TP FP TN FN Sensitivity Specificity
stx1 16 1c 1,227 0 100 (76–100) 99.9 (99–100)
stx2 9 0 1,235 0 100 (63–100) 100 (99–100)
stx1 + stx2 11 0 1,233 0 100 (68–100) 100 (99–100)
a

TP, true positive; FP, false positive, TN, true negative; FN, false negative.

b

CI, confidence interval.

c

Identified as stx1 by the ProGastro SSCS assay and was negative by EIA and bidirectional sequencing analysis.