Table 3.

Complications of ICD implantation procedures for the overall cohort and recipients of replacement or new ICDs.

Event Total
n = 463978
Replacement
ICDS
n = 103985
New ICDs
n = 359993
Absolute
Standardized
Difference (%)
Any adverse event 12453 (2.684%) 945 (0.909%) 11508 (3.197%) 10.3
Cardiac arrest 1225 (0.264%) 101 (0.097%) 1124 (0.312%) 2.0
Drug reaction 365 (0.079%) 50 (0.048%) 315 (0.088%) 0.8
Cardiac perforation 289 (0.062%) 10 (0.010%) 279 (0.078%) 2.1
Coronary venous dissection 511 (0.110%) 32 (0.031%) 479 (0.133%) 2.4
Hematoma 3554 (0.766%) 368 (0.354%) 3186 (0.885%) 4.3
Lead Dislodgement 3959 (0.853%) 197 (0.189%) 3762 (1.045%) 7.5
Hemothorax or Pneumothorax 2081 (0.449%) 138 (0.133%) 1943 (0.540%) 4.8
Transient ischemic attack or stroke 310 (0.067%) 25 (0.024%) 285 (0.079%) 1.4
Myocardial infarction 107 (0.023%) 4 (0.004%) 103 (0.029%) 1.2
Pericardial tamponade 349 (0.075%) 15 (0.014%) 334 (0.093%) 2.4
Infection Related to Device 100 (0.022%) 10 (0.010%) 90 (0.025%) 0.1

Continuous variables compared using Student's T-test.