Skip to main content
. 2013 Oct 18;8(4):454–459. doi: 10.1007/s11552-013-9544-3

Table 1.

Results of cantilever bending and torsion tests for locked and nonlocked groups (mean ± SD)

Construct Bending stiffness (N/mm)*° Maximum bending load (N)*° Torsional stiffness (Nmm/deg)*° Maximum torque (Nmm)*°
Locked, 4 cortices/segment 5.4 ± 0.7 184 ± 27 69.6 ± 9.3 1,858 ± 161
Nonlocked, 6 cortices/segment 5.5 ± 0.5 185 ± 22 68.6 ± 7.9 1,862 ± 196

*p > 0.05, no statistical significant difference; °p < 0.05, statistical equivalence