Skip to main content
. 2013 Oct 23;30(12):1583–1585. doi: 10.1007/s10815-013-0104-0

Table 1.

Temperatures within two Forma (Series II, Model 3110) front-loading incubators. (Data are pooled for both incubators)

Shelf position Front of shelf temperatures (°C; n = 60/location) Back of shelf temperatures (°C; n = 60/location) P value
Top 36.92 ± 0.04 36.65 ± 0.05 < .001
Middle 36.78 ± 0.24 36.64 ± 0.07 < .001
Bottom 36.85 ± 0.05 36.75 ± 0.05 < .001
P value < .001 < .001

Means ± standard deviation