Table 1.
Philips 3T Achieva initial flip angle and absorbed power PA in an RF transducer prototype with reactance X = 211 Ω for various loop resistances (R). The results for a particular human subject (89 kg) are also included. All scans are performed using the same pulse sequence.
Initial flip angle (deg) | ||
---|---|---|
Loop R (Ω) | determined in Achieva | PA(W) |
1000 | 90.3 | 9.8 |
700 | 90.9 | 14.3 |
375 | 87.9 | 25.1 |
Human subject | 90.2 | 24.5 |