. 2013 Nov 15;40(12):122303. doi: 10.1118/1.4829527

Table 3.

Calibration curve linearity and slope data for the final RF dosimeters on 127.8 and 123.2 MHz systems. In brackets is the software version running on each scanner.

  No. of data Slope  
  points (calibration line) R2
3T MRI systems, 127.8 MHz
Philips Achieva XMR Dual-Channel, quadrature mode [R3.2.1] 29 30.02 (0.07 SD) 0.9999
Philips Achieva Dual-Channel, quadrature mode [R3.2.1] 29 29.96 (0.09) 0.9996
Philips Intera [R3.2.2] 29 29.06 (0.04) 0.9998
GE MR750 [22M4] 10 29.82 (0.12) 0.9998
GE HdX (I) [15M4] 11 29.10 (0.13) 0.9999
GE HdX (II) [15M4] 10 28.30 (0.14) 0.9998
All 127.8 MHz combined data 118 29.32 (0.07) 0.9994
3T MRI systems, 123.2 MHz
Siemens Trio (I) [VB17a] 14 27.11 (0.10) 0.9998
Siemens Trio (II) [VB17a] 14 26.82 (0.14) 0.9996