Skip to main content
. Author manuscript; available in PMC: 2013 Dec 1.
Published in final edited form as: J Chromatogr B Analyt Technol Biomed Life Sci. 2010 Jan 1;878(1):10.1016/j.jchromb.2009.11.005. doi: 10.1016/j.jchromb.2009.11.005

Table 1.

LCMSMS parameters, retention times and internal standards (IStd) for buprenorphine (BUP), norbuprenorphine (NBUP), buprenorphine-glucuronide (BUP-Gluc), and norbuprenorphine-glucuronide (NBUP-Gluc).

Analyte MRM
Transitiona
DPb
(V)
EPc
(V)
CEPd
(V)
CEe
(eV)
CXPf
(V)
RTg
(min)
IStd

NBUP-Gluc 590.3>414.3 96 9.5 34 51 4 3.3 NBUP-d3
590.3>396.2 96 9.5 27 50 3

BUP-Gluc 644.3>468.1 101 9.5 30 45 4 4 BUP-d4
644.3>396.3 101 9.5 30 71 4

NBUP 414.3>340.1 86 12 26 43 4 4.2 NBUP-d3
414.3>326 86 12 26 40 5

NBUP-d3 417.3>343.4 86 12 16 39 4 4.2
417.3>326 86 12 22 43 4

BUP 468.3>396.2 86 12 18 53 6 5 BUP-d4
468.3>414.3 86 12 18 45 12

BUP-d4 472.5>400.1 86 11 24 61 6 5
472.5>415.3 86 11 24 45 8
a

The quantification transition is underlined

b

declustering potential

c

entrance potential

d

collision cell entrance potential

e

collision energy

f

collision cell exit potential

g

retention time.