. Author manuscript; available in PMC: 2014 Aug 1.
Published in final edited form as: Circ Arrhythm Electrophysiol. 2013 Jul 24;6(4):10.1161/CIRCEP.113.000114. doi: 10.1161/CIRCEP.113.000114

Table 1.

Performance evaluation of phase estimation during acute ischemia. For each of the body-surface, CS, LV and RV leads the ROC-derived thresholds, determined in artificially induced TWA in control swine, have been applied to the coherence estimation results reported in Figure 3. The 95% confidence intervals are shown in parentheses.

Lead
Surface CS LV RV
Sn 0.89 (0.88–0.90) 0.93 (0.92–0.94) 0.94 (0.93–0.94) 0.80 (0.79–0.81)
Sp 0.93 (0.92–0.94) 0.95 (0.94–0.95) 0.92 (0.91–0.92) 1.00 (1.00-1.00)
PV+ 0.93 (0.92–0.93) 0.95 (0.94–0.95) 0.92 (0.91–0.93) 1.00 (1.00-1.00)
PV− 0.90 (0.89–0.90) 0.93 (0.92–0.94) 0.93 (0.93–0.94) 0.83 (0.83–0.84)
Accuracy 0.91 (0.90–0.92) 0.94 (0.94-0.94) 0.93 (0.92–0.94) 0.90 (0.90-0.90)

(Sn: Sensitivity; Sp: Specificity; PV+: Positive predictive value; PV−: Negative predictive value).