Table 2.

Sensitivity and regression parameters.

Parameter Method B Method C
λ max⁡, nm 460 470
Colour stability, min 60 15
Linear range, μg mL−1 0.4–8.0 0.4–10
Molar absorptivity (ε), L mol−1 cm−1 4.0 × 104 3.7 × 104
Sandell sensitivity*, μg cm−2 0.0105 0.0136
Limit of detection (LOD), μg mL−1 0.39 0.43
Limit of quantification (LOQ), μg mL−1 1.19 1.32
Regression equation, Y**
 Intercept (a) 0.8744 0.8082
 Slope (b) −0.0914 −0.0815
 Standard deviation of a (S a) 2.2 × 10−2 6.6 × 10−2
 Standard deviation of b (S b) 4.9 × 10−3 7.3 × 10−3
 Regression coefficient (r) −0.9998 −0.9990

 *Limit of determination as the weight in μg per mL of solution, which corresponds to an absorbance of A = 0.001 measured in a cuvette of cross-sectional area 1 cm2 and l = 1 cm. **Y = a + bX, where Y is the absorbance, X is concentration in μg mL−1, a is intercept, and b is slope.