. 2013 Sep 10;8:212. doi: 10.1186/1748-717X-8-212

Table 1.

Error components for the presented setup correction protocol in different directions (mm)

Structure Systematic error Σ (1 SD)
Random error σ (1 SD)
AP/ vertical CC/ longitudinal LAT/ lateral AP/ vertical CC/ longitudinal LAT/ lateral
Rigid target1
0.8 (1.1)
0.9 (1.3)
0.7 (0.7)
1.7 (1.3)
1.6 (1.6)
1.6 (1.2)
Observer variation
0.9 (1.0)
0.8 (1.0)
0.9 (0.9)
1.4 (1.4)
1.5 (1.6)
1.6 (1.6)
C1-22
2.0 (1.6)
1.0 (1.4)
1.6 (1.3)
1.8 (1.4)
1.3 (1.6)
1.6 (1.5)
C5-72
1.5 (1.8)
1.0 (1.5)
1.2 (1.2)
1.4 (1.6)
1.8 (1.9)
1.7 (1.6)
Mandible2
2.9 (1.5)
2.7 (1.5)
1.6 (1.3)
2.3 (1.7)
2.4 (1.9)
1.9 (1.6)
Occiput bone2 2.4 (2.2) 1.5 (2.0) - 2.3 (1.9) 1.9 (1.7) -

1Errors of the reference location for weekly IGRT without observer related variation.

2Errors with daily imaging including observer errors.

The results are given for the two matching principles as MID_PTV (MIN_MAX).