Skip to main content
. 2013 Oct 14;8:237. doi: 10.1186/1748-717X-8-237

Table 4.

Comparison of delivery efficiency in all four radiotherapy plan groups

  VMAT1 VMAT2 dIMRT ssIMRT pGLM-u pVMAT1 -VMAT2 pVMAT1 -dIMRT pVMAT1 -ssIMRT pVMAT2 -dIMRT pVMAT2 -ssIMRT pdIMRT -ssIMRT
MUs
1232.1 ± 146.2
1349.9 ± 133.8
1292.2 ± 120.7
1090.2 ± 91.1
<0.001
0.001
0.067
<0.001
0.079
<0.001
<0.001
Time (min)
2.7 ± 0.2
3.9 ± 0.3
5.7 ± 0.2
14.1 ± 1.0
<0.001
<0.001
<0.001
<0.001
<0.001
<0.001
<0.001
Mean dose rate (MU/min)
494.0 ± 38.6
379.6 ± 30.1
249.8 ± 17.9
86.6 ± 5.1
<0.001
<0.001
<0.001
<0.001
<0.001
<0.001
<0.001
Control points
161.4 ± 13.1
187.0 ± 30.9
126.6 ± 10.2
138.0 ± 10.4
<0.001
<0.001
<0.001
<0.001
<0.001
<0.001
0.223
plan calculation time (hour) 5.6 ± 1.4 4.8 ± 1.2 0.8 ± 0.3 0.8 ± 0.4 <0.001 0.092 <0.001 <0.001 <0.001 <0.001 0.945

GLM-u: general linear model-univariate; VMAT1: single arc VMAT, VMAT2: dual arc VMAT; dIMRT: dynamic MLC IMRT; ssIMRT: step-and-shoot IMRT. “Bold numbers” means that p value has a statistically significance.