Table 1.

Radius of curvature of posterior eye in millimeters for OCT images (original and corrected with numerical and analytical techniques) and for a reference technique (MRI).

Subject/Eye Original Numerical
(Ray Traced)
Analytical
(Reduced Eye)
MRI
1/OD 69.30 14.92 16.02 12.04
1/OS 40.66 13.23 13.01 11.46
2/OD 17.54 10.01 8.37 12.24
2/OS 24.15 11.18 8.74 12.97
3/OD 98.31 15.37 21.04 12.42
3/OS 128.95 14.62 21.22 12.92
4/OD 31.55 10.94 11.13 10.61
4/OS 51.38 12.58 12.25 11.02
5/OD 52.51 15.40 14.12 12.30
5/OS 134.13 17.81 13.83 12.32