Reduction of noise and dose using iterative reconstruction and the stellar detector

Soft tissue
Air
Bone
Lowest dose Highest dose Lowest dose Highest dose Lowest dose Highest dose
Δ HUsd Range Δ HUsd Range Δ HUsd Range
Noise reduction from FBP to IR 10 (31%) 15 (30%) to 7 (31%) 31 (44%) 53 (49%) to 22 (41%) 64 (30%) 128 (37%) to 28 (21%)
Noise reduction from IR to IR with Sd 7 (32%) 16 (42%) to 4 (24%) 12 (31%) 19 (34%) to 9 (28%) 24 (15%) 51 (22%) to 9 (7%)
Δ DLP Range Δ DLP Range Δ DLP Range
Dose reduction from FBP to IR for the same noise level 53 (52%) 116 (55%) to 25 (50%) 133 (80%) 165 (78%) to 106 (81%) 48 (30%) 24 (12%) to 40 (55%)
Dose reduction from IR to IR with Sd for the same noise level 69 (54%) 96 (39%) to 46 (65%) 106 (70%) 166 (67%) to 65 (72%) 35 (27%) 33 (21%) to 25 (43%)

HUsd: Noise (sd of Hounsfield Units); FBP: Filtered back projection; IR: Iterative reconstruction; Sd: Stellar detector; DLP: Dose length product.