Table 5.

Parameter for left-sided breast cancer cases, including PTV coverage, VMAT, age and dose to the lungs and heart

Parameters 3D-CRT IMRT VMAT 3D-CRT vs IMRT IMRT vs VMAT 3D-CRT vs VMAT
PTV
 V95% 93.479±2.25% 94.213±2.76% 89.732±3.15% 0.704 0.011 0.094
 D95 (Gy) 45.689±3.68 (91.378%) 47.197±0.88 (94.394%) 46.289±0.71 (92.578%) 0.382 0.02 (<0.05) 0.705
Lung—left
 V40 13.142%±7.76 7.699%±2.12 7.126%±2025 0.09 0.661 0.116
 V20 21.905%±4.97 19.812%±2.67 25.06%±4.50 0.168 0.055 0.274
 V5 34.343%±6.34 60.420%±11.75 93.395%±4.24 0.002 (<0.01) 0.0008 (<0.001) <0.0001 (<0.0001)
 Dmean (Gy) 11.243±2.53 12.297±1.64 16.293±0.96 0.178 0.0096 (<0.01) 0.009 (<0.01)
Lung—right
 V5% 0.000%±0.00 0.148%±0.3 14.796%±12.29 9.276 0.031 (<0.05) 0.032 (<0.05)
 Dmean (Gy) 0.153±0.06 0.697±0.19 5.592±5.94 0.002 (<0.01) 0.1 0.076
Heart
 V40 6.548%±3.85 1.988%±0.60 2.330%±1.50 0.02 (<0.05) 0.0533 0.016
 V20 11.657%±2.50 12.779%±3.54 15.369%±6.14 0.232 0.249 0.126
 V5 18.568%±3.44 46.675%±13.01 85.921%±15.92 0.003 (<0.01) 0.002 (<0.01) 0.0001 (<0.001)
 D5 (Gy) 43.489±4.16 33.475±3.78 33.396±4.82 0.0004 (<0.001) 0.970 0.001 (<0.01)
 Dmean (Gy) 6.557±1.24 8.782±1.49 12.412±2.47 0.004 (<0.01) 0.015 (<0.05) 0.001 (<0.01)

3D, three dimensional; CRT, conformal radiotherapy; D95, the dose distributed in ≥95% of the PTV; Dmean, mean dose; IMRT, intensity-modulated radiotherapy; PTV, planning target volume; V20, volume receiving 20 Gy; V40, volume receiving 40 Gy; V5, volume receiving 5 Gy; V95%, volume receiving 95% of the prescribed dose; VMAT, volumetric modulated arc therapy.