Table 2.

Comparison of neutron source strength for various medical linacs.

Manufacturer Model Energy (MeV) Q (n Gy−1) Study
Siemens KD 20 0.92 × 1012 McCall (1987)2
Siemens Primus 15 0.20 × 1012 Lin et al. (2001)47
Siemens Primus 15 0.17 × 1012 Pena et al. (2005)26
Siemens Primus 15 0.136 × 1012 Becker et al. (2007)13
Siemens MD2 10 0.08 × 1012 Followill et al. (2003)17
Siemens MD 15 0.2 × 1012 Followill et al. (2003)17
Siemens KD 18 0.88 × 1012 Followill et al. (2003)17
Siemens Primus (with MiMIC) 10 0.02 × 1012 Followill et al. (2003)17
Siemens Primus (with MiMIC) 15 0.12 × 1012 Followill et al. (2003)17
Siemens Primus (with MLC) 15 0.21 × 1012 Followill et al. (2003)17
Varian 1800C 18 2.9 × 1012 McCall (1987)2
Varian 20C 15 0.93 × 1012 McCall (1987)2
Varian 18C 10 0.059 × 1012 McCall (1987)2
Varian 1800C 18 2.27 × 1012 McGinley and Landry (1989)56
Varian 1800C 15 1.23 × 1012 McGinley and Landry (1989)56
Varian 1800C 10 0.06 × 1012 McGinley and Landry (1989)56
Varian 2100C/2300C 20 1.2 × 1012 Mao et al. (1997)32
Varian 2100C/2300C 18 1.2 × 1012 Mao et al. (1997)32
Varian 2100C/2300C 15 6.8 × 1011 Mao et al. (1997)32
Varian 2100C/2300C 10 3.8 × 1010 Mao et al. (1997)32
Varian 2100C 18 0.96 × 1012 Followill et al. (2003)17
Varian 2100C (with MLC) 18 0.87 × 1012 Followill et al. (2003)17
Varian 2300 CD 18 0.95 × 1012 Followill et al. (2003)17
Varian 2500 24 0.77 × 1012 Followill et al. (2003)17
GE Saturne 43 25 2.4 × 1012 Fenn and McGinley (1995)58
GE Saturne 43 18 1.5 × 1012 Fenn and McGinley (1995)58
GE Saturne 41 15 0.47 × 1012 Fenn and McGinley (1995)58
GE Saturne 41 12 0.24 × 1012 Fenn and McGinley (1995)58
Elekta SL-20 17 0.69 × 1012 McGinley et al. (1993)57
Elekta SL-25 22 2.37 × 1012 McGinley et al. (1993)57
Elekta SL-20 18 0.46 × 1012 Followill et al. (2003)17
Elekta SL-25 18 0.46 × 1012 Followill et al. (2003)17