Table 2.

The recommended values [R] and measured values [M] for EDW for energy 15 MV.

Orientation of EDW R [°] M [°] Difference Percentage difference [%]
IN 10 10.38 −0.38 3.80
OUT 10 10.34 0.34 3.40
IN 15 15.06 −0.06 0.40
OUT 15 15.22 0.28 1.50
IN 20 19.35 0.65 −3.30
OUT 20 20.02 0.02 0.10
IN 25 24.7 0.30 −1.20
OUT 25 24.91 0.90 −0.40
IN 30 29.45 0.55 −1.83
OUT 30 29.71 0.29 −1.00
IN 45 43.94 1.06 −2.36
OUT 45 44.31 0.69 −1.50
IN 60 58.95 1.05 −1.75
OUT 60 59.03 0.97 −1.60