Table 2.
The recommended values [R] and measured values [M] for EDW for energy 15 MV.
| Orientation of EDW | R [°] | M [°] | Difference | Percentage difference [%] |
|---|---|---|---|---|
| IN | 10 | 10.38 | −0.38 | 3.80 |
| OUT | 10 | 10.34 | 0.34 | 3.40 |
| IN | 15 | 15.06 | −0.06 | 0.40 |
| OUT | 15 | 15.22 | 0.28 | 1.50 |
| IN | 20 | 19.35 | 0.65 | −3.30 |
| OUT | 20 | 20.02 | 0.02 | 0.10 |
| IN | 25 | 24.7 | 0.30 | −1.20 |
| OUT | 25 | 24.91 | 0.90 | −0.40 |
| IN | 30 | 29.45 | 0.55 | −1.83 |
| OUT | 30 | 29.71 | 0.29 | −1.00 |
| IN | 45 | 43.94 | 1.06 | −2.36 |
| OUT | 45 | 44.31 | 0.69 | −1.50 |
| IN | 60 | 58.95 | 1.05 | −1.75 |
| OUT | 60 | 59.03 | 0.97 | −1.60 |