Skip to main content
. 2011 Aug 15;17(1):4–12. doi: 10.1016/j.rpor.2011.06.007

Table 2.

Comparison of intensity maps of physical and enhanced dynamic wedges for 10 cm × 10 cm field size at the depth of dmax, 10 cm and 20 cm for 6 and 18 MV photons.

Wedge angle DD
DTA
γ ≤ 1
DD
DTA
γ ≤ 1
DD
DTA
γ ≤ 1
dmax 10 cm 20 cm
6 MV photons
 15° 2% 2 mm 100% 2% 2 mm 100% 2% 2 mm 100%
 30° 3% 2 mm 100% 2% 2 mm 100% 3% 3 mm 100%
 45° 3% 3 mm 98% 3% 3 mm 98% 3% 3 mm 97.4%
 60° 3% 3 mm 96.6% 3% 3 mm 96.7% 3% 3 mm 100%
18 MV photons
 15° 2% 2 mm 100% 2% 2 mm 100% 2% 2 mm 100%
 30° 2% 2 mm 100% 2% 2 mm 100% 2% 2 mm 100%
 45° 2% 2 mm 100% 2% 2 mm 100% 2% 2 mm 100%
 60° 3% 3 mm 97.6% 3% 3 mm 97.2% 3% 3 mm 100%

DD: delta dose, DTA: distance to agreement, γ ≤ 1: gamma pixel match.