Table 2.

Three verification measurements [Gy] for the 4 MeV electron beam, their mean value and the percentage difference between the mean value and the prescribed dose of 3 Gy.

Applicator diameter [cm] d1 d2 d3 Mean Difference [%]
3.0 3.323 3.332 3.339 3.331 11.04
3.5 3.257 3.265 3.263 3.262 8.72
4.0 3.283 3.287 3.303 3.291 9.70
4.5 3.255 3.256 3.260 3.257 8.57
5.0 3.205 3.227 3.223 3.218 7.28
5.5 3.283 3.286 3.293 3.287 9.58
6.0 3.192 3.206 3.191 3.196 6.54
6.5 3.109 3.119 3.151 3.126 4.21
7.0 3.154 3.160 3.156 3.157 5.22
7.5 3.132 3.135 3.131 3.133 4.42
8.0 3.093 3.089 3.091 3.091 3.03
8.5 3.098 3.098 3.103 3.100 3.32
9.0 3.100 3.102 3.100 3.101 3.36
9.5 3.081 3.084 3.085 3.083 2.78
10.0 3.064 3.058 3.062 3.061 2.04