Skip to main content
. 2013 Feb 4;18(4):201–208. doi: 10.1016/j.rpor.2012.12.005

Table 1.

Values of planning parameters used to create eight different treatment plans. FW – field width, PF – pitch factor, and MF – modulation factor.

Plan name FW PF MF planned MF actual PTV1
Mean ± SD
MF actual PTV2
Mean ± SD
Reference 1.05 0.107 3.5 2.47 ± 0.32 2.26 ± 0.39
Default 2.5 0.215 2.5 2.01 ± 0.14 1.94 ± 0.22
FW = 1.05 1.05 0.215 2.5 2.05 ± 0.28 2.17 ± 0.14
FW = 5 5 0.215 2.5 1.86 ± 0.14 1.88 ± 0.13
PF = 0.107 2.5 0.107 2.5 1.89 ± 0.16 1.77 ± 0.15
PF = 0.43 2.5 0.43 2.5 2.10 ± 0.13 2.02 ± 0.16
MF = 1.5 2.5 0.215 1.5 1.25 ± 0.06 1.31 ± 0.05
MF = 3.5 2.5 0.215 3.5 2.80 ± 0.21 2.64 ± 0.22