Skip to main content
. Author manuscript; available in PMC: 2015 Jan 1.
Published in final edited form as: J Urol. 2013 Aug 2;191(1):10.1016/j.juro.2013.07.087. doi: 10.1016/j.juro.2013.07.087

Table 2.

Hemorrhagic lesions of kidneys exposed to ultrasonic propulsion with clinical prototype, high power or continuous duty cycle using research ultrasound machine, or SWL.

Transcutaneous Probe on Kidney SWL

Ultrasonic Propulsion High Power High Power High Duty Cycle
No. kidneys 3 5 3 3 6
Machine Clinical Research Research Research Dornier HM-3
Stone/bead Yes No No No No
Power 10 W 240 W 240 W 80 W 24 kV
Duty cycle 3.3% 3.3% 3.3% 100% 120 SWs/min
Time (mins) 1.7±0.1 10 10 10 20
p+ (MPa) 12 68 93 47 40–55
SPPA inten. (W/cm2) 2,400 15,000 26,130 9,320

% lesion size 0.0±0.0 0.0±0.0 0.46±0.37 1.15±0.49 1.56±0.45