Table 1.
Table of values for the linear regression of the nanowire response to Mouse IgG vs. the gate oxide thickness from Figure 5B. The limit of detection was taken at the 95% confidence level.
TOX(A˚́) | Slope(mV/log (CMouse IgG )) | Mouse IgG at VT = 0(pg/mL) | LOD 95% C.I. (pg/mL) |
---|---|---|---|
50 | −302.0 | 0.0383 | 0.0536 |
90 | −180.1 | 0.0477 | 0.125 |
150 | −32.5 | 0.0578 | 0.150 |