. 2013 Dec 19;3:311. doi: 10.3389/fonc.2013.00311

Table 2.

Some of the discrepancies detected during RPC dosimetry review visits to 156 institutions in 2005–2011.

Errors regarding Number of institutions (%)
Inadequacy of QA program 115 (74)
Photon field-size dependence (small fields)* 62 (40)
Wedge transmission factor* 50 (32)
Off-axis factors, beam symmetry 46 (29)
Electron calibration* 27 (17)
Photon depth dose* 25 (16)
Electron depth dose* 18 (11)
Photon calibration* 13 (8)

The parameters indicated by “*” are considered significant dosimetry parameters that influence the calculation of patient treatment meter settings. During this 6-year period, 70% of the visited institutions received one or more recommendations to address discrepancies in these dosimetry parameters.