Skip to main content
. 2013 Oct 24;13(11):14367–14397. doi: 10.3390/s131114367

Table 7.

Mass in the liquid of the medium accuracy hydrometer. The mL values include the air pressure corrections and the calibrated standards. It includes the standard deviation σmL of the mL measurements.

Hydrometer 08482 (Medium Accuracy) Hydrometer 08482 (Medium Accuracy)


Traditional Method Mass in the Liquid at Point 1,025 kg/m3 Method Using the Vision System Mass in the Liquid at Point 1,025 kg/m3
B Hydrometer A Standard B Hydrometer A Standard
l1 15.4332g 15.0001g l1 15.4367g 15.0003g
l2 15.4425g 15.0000g l2 15.4368g 15.0002g
l3 15.4317g 15.0001g l3 15.4367g 15.0004g
l4 15.4268g 15.0001g l4 15.4366g 15.0003g
l5 15.4219g 15.0002g l5 15.4367g 15.0003g
l6 15.4440g 15.0003g l6 15.4368g 15.0004g
mL 15.43122g 15431.22mg mL 15. 43459g 15434.59mg
σ mL 0.00868g 8.68mg σ mL 0.00012g 0.12mg