Table 3.
Memory Metal Spinal System | |||
---|---|---|---|
Specimen |
Load (N) |
Cycles to failure (n) |
Failure mode |
1 |
198 |
5000000 |
No observed failure |
2 |
237 |
5000000 |
No observed failure |
3 |
237 |
5000000 |
No observed failure |
4 |
269 |
771901 |
Superior screws rotate inferiorly about superior transverse bridge |
5 |
296 |
500 |
Superior screws rotate inferiorly about superior transverse bridge |
6 |
355 |
250 |
Superior screws rotate inferiorly about superior transverse bridge |
MOSS Miami | |||
Specimen |
Load (N) |
Cycles to failure (n) |
Failure mode |
1 |
160 |
5000000 |
No observed failure |
2 |
170 |
5000000 |
No observed failure |
3 |
172 |
80000 |
Screw head/shank interface |
4 |
175 |
134537 |
Screw head/shank interface |
5 |
180 |
90811 |
Screw head/shank interface |
6 | 220 | 44251 | Screw head/shank interface |