TABLE 2.
Sample no. | Mean no. of HIV-1 RNA copies/ml detected by the following protocol:
|
SD
|
Coefficient of variation (%)
|
|||
---|---|---|---|---|---|---|
Automateda | Manualb | Automated | Manual | Automated | Manual | |
1 | 8.7 × 101 | 1.1 × 102 | 5.8 × 101 | 8.0 × 101 | 66 | 69 |
2 | 4.3 × 103 | 6.2 × 103 | 2.3 × 103 | 2.2 × 103 | 53 | 36 |
3 | 8.0 × 103 | 1.3 × 104 | 2.9 × 103 | 5.9 × 103 | 36 | 46 |
4 | 9.8 × 103 | 5.7 × 103 | 2.8 × 103 | 2.4 × 103 | 28 | 41 |
5 | 1.9 × 104 | 1.5 × 104 | 6.2 × 103 | 3.7 × 103 | 32 | 24 |
6 | 2.0 × 104 | 2.0 × 104 | 8.9 × 103 | 1.9 × 103 | 44 | 10 |
7 | 2.8 × 104 | 2.9 × 104 | 8.1 × 103 | 7.9 × 103 | 29 | 27 |
8 | 6.6 × 104 | 5.3 × 104 | 1.4 × 103 | 2.6 × 104 | 21 | 48 |
Automated sample preparation followed by the LCx HIV RNA Quantitative assay.
Manual sample preparation followed by the LCx HIV RNA Quantitative assay.