. 2004 Apr;42(4):1439–1443. doi: 10.1128/JCM.42.4.1439-1443.2004

TABLE 3.

Intra-assay variation of results obtained by the automated versus the manual sample preparation protocola

Sample no. Mean no. of HIV-1 RNA copies/ml detected by the following protocol:
SD
Coefficient of variation (%)
Automatedb Manualc Automated Manual Automated Manual
1 9.7 × 102 9.3 × 102 2.1 × 102 1.7 × 102 22 18
2 7.3 × 103 5.1 × 103 1.8 × 103 3.6 × 102 25 7
3 3.5 × 104 1.6 × 104 8.4 × 103 2.6 × 103 24 17
4 5.1 × 104 3.6 × 104 3.6 × 103 6.8 × 103 7 19
a

Data are from five replicates of one run.

b

Automated sample preparation followed by the LCx HIV RNA Quantitative assay.

c

Manual sample preparation followed by the LCx HIV RNA Quantitative assay.