TABLE 1.

Results by samplea

Virus ENIVD sample Strainb Copies/ml expected Copies/ml determinedc % Of laboratories achieving positive resultsd
Ebola F6 Mayinga e 2.2E7 92.9
F9 Mayinga 2.2E6 2.7E6 85.7
F4 Mayinga 2.2E5 1.6E5 78.6
Marburg F5 Musoke 3.6E7 78.6
F3 Ratayczak 3.9E4 64.3
F10 Ratayczak 3.9E3 3.9E3 57.1
F18 Ratayczak 3.9E2 <2,647f 35.7
Lassa L11 AV 4.0E5 5.4E5 85.7
L4 AV 4.0E4 2.6E4 78.6
L8 AV 4.0E3 2.6E3 50
L17 AV 4.0E2 <2,445f 21.4
L6 Josiah 4.8E6 3.8E6 71.4
L9 Josiah 4.8E5 7.0E5 85.7
L3 Josiah 4.8E4 1.2E5 71.4
L2 Josiah 4.8E3 7.2E3 50
Orthopox P12 MPX Lam87 5.0E7 3.2E7 100
P1 MPX Lam87 5.0E6 3.4E6 95.6
P10 MPX Lam87 5.0E5 5.6E5 81.8
P5 MPX Lam87 5.0E4 8.0E4 68.2
P14 MPX Lam87 5.0E3 8.0E3 59.1
P3 MPX Lam87 5.0E2 1.6E3 45.5
P6 MPX Lam87 5.0E1 <1,000f 27.3
P7 CML 8.8E6 8.8E6 95.6
P9 CPXV 81/02 6.0E5 86.4
P15 CPX Brithon 2.4E6 95.6
P13 VAC 1.0E6 90.9
P11 VAC mod.g 1.6E5 86.4
P4 ECT 3.6E6 72.7
a

All samples available on request through ENIVD (www.enivd.de).

b

Orthopox virus species: monkeypox virus (MPX), vaccinia virus (VAC), cowpox virus (CPX), camelpox virus (CML), mousepox virus (ectromelia virus [ECT]).

c

Measured by real-time PCR to exclude degradation of virus in plasma after heating and irradiation.

d

Coefficients of correlation of detection rates with virus concentration in samples: filovirus panel, r = 0.914; Lassa virus panel, r = 0.85; and orthopox virus panel, r = 0.914 (P values for the correlations were 0.004, 0.0081, and ≤0.00009, respectively). Positive test results for the total panel for Ebola and Marburg virus, Lassa virus, and orthopox virus, 70, 73, and 78%, respectively; positive test results for samples with ≤104 virus copies, 46, 40, and 44%, respectively.

e

—, virus concentration originally determined in this sample.

f

Virus detected but below linear quantification range of assay.

g

mod., modified.